FDA Adverse Event Other Summary report: N

ACRYSOF RESTOR

MDR report key: 2095708 · Received May 6, 2011

Report

Report Number
1119421-2011-00541
Event Type
Other
Date Received
May 6, 2011
Date of Event
January 1, 2011
Report Date
April 7, 2011
Manufacturer
ALCON RESEARCH, LTD. / HUNTINGTON
Product Code
MFK
PMA / PMN Number
P040020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. RESULTS FROM THE PRODUCT HISTORY RECORD REVIEW INDICATED THE PRODUCT MET RELEASE CRITERIA. ROOT CAUSE: THE ROOT CAUSE COULD NOT BE IDENTIFIED BY THE INVESTIGATION. THE SAMPLE WAS NOT RETURNED. ACTION TAKEN: INVESTIGATION HAS BEEN COMPLETED BASED ON CURRENT INFORMATION. CONCLUSION: BASED ON OUR CURRENT TRACKING, THERE ARE NO ADVERSE TRENDS FOR THIS REPORT COMPLAINT AND BASED ON THESE FINDINGS THE RESIDUAL RISK REMAINS UNCHANGED AND AT AN ACCEPTABLE LEVEL. THERE HAVE BEEN NO OTHER COMPLAINTS REPORTED IN THE LOT NUMBER. ADDITIONAL INFORMATION WAS REQUESTED ON 04/13/2101 AND 04/28/2011 BY PHONE, FAX, AND MAIL. A COMPLETED QUESTIONNAIRE HAS NOT BEEN RECEIVED. (B)(4).

Description of Event or Problem · 1

A SURGEON REPORTED A PATIENT WITH A REFRACTIVE SURPRISE FOLLOWING INTRAOCULAR LENS (IOL) IMPLANT SURGERY. THE SURGEON REPORTED HE WAS UNSURE OF THE CAUSE OF THE REFRACTIVE SURPRISE, BUT IT COULD BE RELATED TO HIM PERFORMING A CAPSULORRHEXIS OF 6.0 MM, WHICH COULD HAVE POTENTIALLY SHIFTED THE EFFECTIVE LENS POSITION OF THE IOL, CAUSING THE MYOPIC SHIFT. THE SURGEON DOES NOT BLAME THE LENS FOR THE UNEXPECTED OUTCOME. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACRYSOF RESTOR INTRAOCULAR LENS MFK ALCON RESEARCH, LTD. / HUNTINGTON SN6AD1 11071383

Patients

Seq Age Sex Outcome Treatment
1 Other