I-STAT CREATININE CARTRIDGE
Report
- Report Number
- 2245578-2024-00217
- Event Type
- Malfunction
- Date Received
- December 17, 2024
- Date of Event
- December 10, 2024
- Report Date
- January 15, 2025
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- CGL
- PMA / PMN Number
- K973292
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 09-JAN-2025. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED AND RETURNED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AN, PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR CREA CARTRIDGE LOT A24301.
APOC INCIDENT # (B)(4) APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
NA.
ON 10-DEC-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CREA CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT CREATININE RESULTS ON A 48 YEAR OLD MALE SEEN FOR A CT EXAM. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. METHOD DATE COLLECTED TESTED RESULT-MG/DL SAMPLE I-STAT (B)(6) 2024 09:02 IMMEDIATELY 2.8 WB I-STAT (B)(6) 2024 09:11 IMMEDIATELY 2.5 WB I-STAT (B)(6) 2024 09:26 IMMEDIATELY 2.5 WB LAB (B)(6) 2024 09:38 NI 1.37 PLASMA THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1720126 | I-STAT CREATININE CARTRIDGE | CREA CARTRIDGE | CGL | ABBOTT POINT OF CARE INC. | NA | A24301 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Male |