BD PHOENIX¿ NMIC-306
Report
- Report Number
- 1119779-2024-01003
- Event Type
- Malfunction
- Date Received
- December 17, 2024
- Date of Event
- December 11, 2024
- Report Date
- March 26, 2025
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LON
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR HIGH MIC MEROPENEM (MEM) WHEN USING PHOENIX PANEL NMIC-306 (CATALOG NUMBER 449292) BATCH NUMBER UNKNOWN. THE CUSTOMER DID NOT RETURN ISOLATES, PANEL RETURNS OR PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE BATCH NUMBER WAS NOT PROVIDED, THEREFORE THIS COMPLAINT IS UNCONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS COULD NOT BE PERFORMED SINCE A BATCH NUMBER WAS NOT PROVIDED. COMPLAINT TRENDING WAS PERFORMED AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.
UNKNOWN LOT NUMBER: D.4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H.4. DEVICE MANUFACTURE DATE: UNKNOWN. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: G4. PMA / 510(K)#: K020322, K022129, K023444, K023634, K023858, K024153, K031530, K031699, K031912, K032299, K032567, K032655, K033362, K033560, K041384, K042932, K052269, K060214, K060217, K060257, K060444, K060447, K061327, K061355, K062207, K062944, K063301, K063486, K063573, K063811, K063824, K071623, K132674, K132909, K151320, K181665, K173252, K190905, K163637, K173523, K033458, AND K123266. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC-306 AN UNSPECIFIED NUMBER OF PATIENT SAMPLES HAVE HIGH MICS FOR THE DRUG MEROPENEM, BUT THE E-TEST WAS SENSITIVE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC-306 AN UNSPECIFIED NUMBER OF PATIENT SAMPLES HAVE HIGH MICS FOR THE DRUG MEROPENEM, BUT THE E-TEST WAS SENSITIVE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1487320 | BD PHOENIX¿ NMIC-306 | SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION | LON | BECTON DICKINSON & CO. (SPARKS) | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |