FDA Adverse Event Malfunction Summary report: N

SERVO-S BASE UNIT

MDR report key: 20941237 · Received December 16, 2024

Report

Report Number
8010042-2024-0002057
Event Type
Malfunction
Date Received
December 16, 2024
Date of Event
December 7, 2024
Report Date
December 16, 2024
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K123149
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON-SITE INVESTIGATION WAS PERFORMED BY FIELD SERVICE ENGINEER. THE CLAIMED ISSUE WAS REPRODUCED DURING TROUBLESHOOTING. ACCORDING TO RECEIVED INFORMATION IN SERVICE REPORT, REPLACEMENT OF THE PRESSURE TRANSDUCER PRINTED CIRCUIT BOARDS (PCB) SOLVED THE ISSUE. THE VENTILATOR PASSED ALL FUNCTIONAL AND SAFETY TESTS AND WAS CLEARED FOR CLINICAL USE. PRESSURE TRANSDUCER PCB MEASURES THE PRESSURE, CONVEYED VIA THE PRESSURE TUBE CONNECTED TO THIS BLOCK BY ITS DIFFERENTIAL PRESSURE TRANSDUCER. WITH DIFFERENTIAL REFERENCE TO THE AMBIENT PRESSURE, THE OUTPUT SIGNAL IS PROPORTIONAL TO THE MEASURED PRESSURE THUS GIVING A LINEAR MEASUREMENT IN THE RANGE -40 CMH2O TO +160 CMH2O. THE DEFECTIVE PARTS WERE NOT RETURNED FOR INVESTIGATION, DESPITE REQUEST. THE DEVICE'S LOGS WERE NOT PROVIDED FOR FURTHER ANALYSIS. THE CAUSE OF THE CLAIMED ISSUE IN THIS CUSTOMER PRODUCT COMPLAINT HAS BEEN DETERMINED. THE ISSUE WAS RELATED TO PRESSURE TRANSDUCER PCB.

Additional Manufacturer Narrative · 0

THE UDI INFORMATION IS NOT AVAILABLE SINCE THE DEVICE WAS MANUFACTURED BEFORE 2015.

Description of Event or Problem · 0

MANUFACTURER'S REF. #: 1183204.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE VENTILATOR FAILED PRESSURE TRANSDUCER TEST DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REF. #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2313074 SERVO-S BASE UNIT VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB 6640440

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown