FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2094022 · Received April 20, 2011

Report

Report Number
1824206-2011-02305
Event Type
Malfunction
Date Received
April 20, 2011
Date of Event
March 28, 2011
Report Date
March 28, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN ISOLATED THE ISSUE TO THE CPR LINKAGE ADJUSTMENTS. HE PROPERLY ADJUSTED THE CPR LINKAGES TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE CPR WAS NOT WORKING TO LOWER THE HEAD SECTION DOWN BUT THE HEAD DOWN FUNCTION WORKED ELECTRICALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1