FDA Adverse Event Malfunction Summary report: N

BD SAFETYGLIDE

MDR report key: 20939199 · Received December 16, 2024

Report

Report Number
1213809-2024-00952
Event Type
Malfunction
Date Received
December 16, 2024
Date of Event
November 13, 2024
Report Date
January 28, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMI
UDI-DI
30382903059172
PMA / PMN Number
K951254
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON PHOTO EVALUATION, IT WAS OBSERVED THAT THE TOP PART OF THE NEEDLE WENT THROUGH THE PLASTIC SHIELD. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED IS CONFIRMED. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 4143704. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE WENT THROUGH THE SHIELD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER TRANSLATED FROM CHINESE TO ENGLISH: PACKAGED DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2327907 BD SAFETYGLIDE NEEDLE, HYPODERMIC, SINGLE LUMEN FMI BECTON DICKINSON MEDICAL SYSTEMS 4143704 30382903059172

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown