FDA Adverse Event Malfunction Summary report: N

B/F GLENOID STRAIGHT SPOKE REAMER ASSEMBLY

MDR report key: 2093905 · Received April 19, 2011

Report

Report Number
1822565-2011-00969
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
February 15, 2011
Report Date
March 21, 2011
Manufacturer
ZIMMER INC
Product Code
HTO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INFORMATION WAS RECEIVED FROM A DISTRIBUTOR WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. AS RETURNED, THE REAMER SHAFT HAS BROKEN AND REMAINS IN THE CANNULATED STRAIGHT DRIVER. THE REAMER HAS A POTENTIAL FIELD AGE OF APPROXIMATELY 1 YEAR. THIS TYPE OF FAILURE MAY OCCUR IF THE THREADED PEG OF THE REAMER IS NOT COMPLETELY AND SECURELY SCREWED INTO THE DRIVER. THIS SITUATION MAY ALLOW THE REAMING TORQUE AND ANY BENDING FORCES TO BE APPLIED TO THE THREADED PEG INSTEAD OF BEING TRANSMITTED BY THE REAMER BODY'S FACETS TO THE TABS OF THE DRIVER. HOWEVER, THE EXACT CAUSE OF FAILURE CANNOT BE DETERMINED WITH CERTAINTY. DEVICE HISTORY RECORDS INDICATE ALL COMPONENTS WERE MANUFACTURED AND INSPECTED TO SPECIFICATION. DIMENSIONAL ANALYSIS FOUND THE PART TO BE WITHIN SPECIFICATION WHERE MEASURED. NO MANUFACTURING ABNORMALITIES COULD BE DETECTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PIECE OF THE REAMER BROKE OFF IN THE DRIVER DURING USE.