FDA Adverse Event Malfunction Summary report: N

ENDEAVOR RESOLUTE RX CORONARY STENT SYSTEM

MDR report key: 2093884 · Received April 6, 2011

Report

Report Number
9612164-2011-00153
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
February 28, 2011
Report Date
March 8, 2011
Manufacturer
MEDTRONIC IRELAND
Product Code
NIQ
PMA / PMN Number
P060033
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): RESULTS: (LESION MORPHOLOGY), (FAILURE TO DELIVER STENT AND STENT DEFORMATION), (USE OF FORCE TO ADVANCE THE STENT). CONCLUSION: (LESION MORPHOLOGY), (USE OF FORCE TO ADVANCE THE STENT). EVALUATION SUMMARY: A NUMBER OF STRUTS ON THE 17TH AND 18TH PROXIMAL STENT SEGMENT WERE PARTIALLY RAISED AND DEFORMED THE DISTAL TIP WAS SLIGHTLY FLARED.

Description of Event or Problem · 1

THE PHYSICIAN WAS ATTEMPTING TO IMPLANT A 3.0 MM DIAMETER X 30MM LENGTH ENDEAVOR RESOLUTE RAPID EXCHANGE (RX) DRUG ELUTING IN A PATIENT. IT WAS REPORTED THAT THE PATIENT HAD A LESION EXHIBITING 95% STENOSIS IN THE MID LCX AND A LESION EXHIBITING 99% STENOSIS IN THE DISTAL RCA, HOWEVER IT IS UNCLEAR WHICH LESION THE PHYSICIAN ATTEMPTED TO TREAT USING THE RESOLUTE STENT. THE DEVICE WAS INSPECTED PRIOR TO USE WITH NO ISSUES NOTED. IT WAS REPORTED THAT THE LESION WAS PRE-DILATED. INFORMATION RECEIVED CONFIRMED THAT RESISTANCE WAS ENCOUNTERED AND FORCE WAS APPLIED IN AN ATTEMPT TO DELIVER THE STENT ACROSS THE LESION. THE PHYSICIAN COULD NOT POSITION THE STENT. AFTER TWO ATTEMPTS THE PATIENT EXPERIENCED CHEST TIGHTNESS AND THE PHYSICIAN ABANDONED THE SURGERY. THE PATIENT FELT WELL ONE HOUR LATER. PLEASE NOTE THAT THIS DEVICE (B)(4) IS DISTRIBUTED OUTSIDE THE UNITED STATES; HOWEVER, IT IS SIMILAR TO THE UNITED STATES DISTRIBUTED PRODUCT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDEAVOR RESOLUTE RX CORONARY STENT SYSTEM NIQ MEDTRONIC IRELAND NA 0003695486

Patients

Seq Age Sex Outcome Treatment
1 UNK