FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2093846 · Received April 6, 2011

Report

Report Number
8020893-2011-00160
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 30, 2011
Report Date
March 30, 2011
Manufacturer
PURITAN BENNETT CORP.
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE PURITAN BENNETT CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BDU PCB. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK PURITAN BENNETT CORP. 840

Patients

Seq Age Sex Outcome Treatment
1