FDA Adverse Event Injury Summary report: N

ASR ACETABULAR CUPS 54

MDR report key: 2093810 · Received May 12, 2011

Report

Report Number
1818910-2011-08184
Event Type
Injury
Date Received
May 12, 2011
Date of Event
January 24, 2011
Report Date
January 10, 2014
Manufacturer
DEPUY INTERNATIONAL, LTD.
Product Code
KWA
PMA / PMN Number
K040627
Removal / Correction Number
Z-1749/1816-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

CORRECTED:CORRECTION/REMOVAL REPORTING NUMBER.THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN (B)(4) 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4).DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.

Description of Event or Problem · 1

PT CONTACTED DEPUY AS A RESULT OF THE ASR RECALL TO INITIATE A CLAIM. MEDICAL RECORDS WERE OBTAINED. MEDICAL RECORDS INDICATE THAT THE PT WAS REVISED TO ADDRESS INCREASING PAIN AND ELEVATED METAL ION LEVELS. INTRAOPERATIVELY FOUND AN ABUNDANT AMOUNT OF CLOUDY FLUID IN THE JOINT. THE CUP DID NOT DISPLAY SOLID BONY INGROWTH.

Description of Event or Problem · 1

UPDATE REC'D 1/10/2014 - LEGAL CLAIM WAS RECEIVED, WHICH IDENTIFIED DOI INFORMATION. THERE WAS NO NEW INFORMATION THAT WOULD CHANGE THE OUTCOME OF THE INVESTIGATION. THE COMPLAINT WAS UPDATED ON: 08/13/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASR ACETABULAR CUPS 54 87KWA KWA DEPUY INTERNATIONAL, LTD. NA 2175994

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention