FDA Adverse Event Injury Summary report: N

FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER

MDR report key: 2093804 · Received May 18, 2011

Report

Report Number
3005099803-2011-01574
Event Type
Injury
Date Received
May 18, 2011
Date of Event
April 22, 2011
Report Date
April 22, 2011
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
GEX
PMA / PMN Number
K100078
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL EXAMINATION OF THE RETURNED DEVICE REVEALED APPROXIMATELY 3 MM OF THE DISTAL TIP WAS RECEIVED SEPARATELY. HANDLING DAMAGE CONTRIBUTED TO THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT DURING A PROCEDURE USING A FLEXIVA 1000 LASER FIBER, THE TIP OF THE LASER FIBER DETACHED INSIDE THE PATIENT FOLLOWING 2 TO 3 FIRES OF THE LASER. THE PHYSICIAN REPORTED THAT ONCE HE BEGAN FIRING THE LASER, SMALL FRAGMENTS OF GLASS DETACHED AT THE BLUE CLADDING OF THE FIBER. THE PHYSICIAN RETRIEVED THE TIP OF THE FIBER WITH GRASPING FORCEPS. LASER ENERGY FROM A 100 WATT LUMENIS LASER SET AT .6 -1 JOULES RATE WAS BEING USED WITH THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT DURING A PROCEDURE USING A FLEXIVA 1000 LASER FIBER, THE TIP OF THE LASER FIBER DETACHED INSIDE THE PATIENT FOLLOWING 2 TO 3 FIRES OF THE LASER. THE PHYSICIAN REPORTED THAT ONCE HE BEGAN FIRING THE LASER, SMALL FRAGMENTS OF GLASS DETACHED AT THE BLUE CLADDING OF THE FIBER. THE PHYSICIAN RETRIEVED THE TIP OF THE FIBER WITH GRASPING FORCEPS. LASER ENERGY FROM A 100 WATT LUMENIS LASER SET AT .6 -1 JOULES RATE WAS BEING USED WITH THE FIBER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXIVA (TM) HIGH POWER SINGLE-USE LASER FIBER POWERED LASER SURGICAL INSTRUMENT GEX BOSTON SCIENTIFIC - MARLBOROUGH M0068403941 1ML0121304

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention 100 WATT LUMENIS LASER