ALINITY C PROCESSING MODULE
Report
- Report Number
- 3016438761-2024-00726
- Event Type
- Malfunction
- Date Received
- December 16, 2024
- Date of Event
- November 26, 2024
- Report Date
- January 13, 2025
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- JJE
- UDI-DI
- 00380740137380
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ABBOTT FIELD SERVICE (FS) INVESTIGATED THE ISSUE ON SITE. FS TROUBLESHOOTING INCLUDED REPLACING THE ALINITY C-SERIES SAMPLE PROBE WHICH RESOLVED THE ISSUE. THERE HAVE BEEN NO FURTHER REPORTS OF DISCREPANT RESULTS SINCE THE PART WAS REPLACED. A REVIEW OF THE ALINITY C PROCESSING MODULE SN# (B)(6) SERVICE HISTORY WAS NOT ABLE TO IDENTIFY OR CONFIRM ANY ADDITIONAL LIKELY CAUSES. A REVIEW OF HISTORICAL DATA REVEALED NO TRENDS, SYSTEMIC ISSUES, OR RELATED NONCONFORMANCES. A REVIEW OF THE CLINICAL CHEMISTRY SYSTEMS REVEALED FOUND NO SYSTEMIC ISSUES OR TRENDS FOR THE ISSUE ASSOCIATED WITH THIS TICKET (ERRATIC/DISCREPANT RESULTS). A REVIEW OF LABELLING ADEQUATELY ADDRESSES SAMPLE RESULTS OBSERVED PROBLEMS FOR ERRATIC / DISCREPANT RESULTS. BASED ON THE INVESTIGATION NO SYSTEMIC ISSUE OR PRODUCT DEFICIENCY WAS IDENTIFIED FOR THE ALINITY C PROCESSING MODULE SN# (B)(6) .
AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
THE CUSTOMER REPORTED A FALSELY DECREASED SODIUM RESULT GENERATED ON THE ALINITY C PROCESSING MODULE ON A PATIENT THAT WAS NOT REPORTED OUT OF THE LABORATORY. RESULTS PROVIDED: SODIUM = 117 / 140 MMOL/L (NORMAL RANGE: 132-146 MMOL/L) NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER REPORTED A FALSELY DECREASED SODIUM RESULT GENERATED ON THE ALINITY C PROCESSING MODULE ON A PATIENT THAT WAS NOT REPORTED OUT OF THE LABORATORY. RESULTS PROVIDED: SODIUM = 117 / 140 MMOL/L (NORMAL RANGE: 132-146 MMOL/L). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2374943 | ALINITY C PROCESSING MODULE | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | ABBOTT LABORATORIES | 00380740137380 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ALNTY C ICT SAMPLE DILU, 07P53-20, 04896UN24| ALNTY C ICT SAMPLE DILU, 07P53-20, 04896UN24 |