FDA Adverse Event Injury Summary report: N

DREAMSTATION AUTO CPAP

MDR report key: 20936319 · Received December 16, 2024

Report

Report Number
2518422-2024-074460
Event Type
Injury
Date Received
December 16, 2024
Date of Event
August 7, 2023
Report Date
December 23, 2024
Manufacturer
RESPIRONICS, INC.
Product Code
BZD
PMA / PMN Number
K131982
Removal / Correction Number
Z-1974-2021
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

IN PREVIOUS INITIAL/FINAL REPORT, 510K NUMBER WAS WRONGLY GIVEN AS K091319 IN THIS REPORT IT IS CORRECTED.

Description of Event or Problem · 0

THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING LUNG DISEASE. THERE WAS NO REPORT OF MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION CAN BE REQUESTED AT THIS TIME. THE MANUFACTURER WAS MADE AWARE OF THIS COMPLAINT THROUGH A REPRESENTATIVE OF THE CUSTOMER. THE DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER FOR FURTHER EVALUATION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND NO EVIDENCE OF FOAM DEGRADATION. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WAS NO ERROR CODE FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY COULDN'T CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS NO VISIBLE FOAM DEGRADATION AND UNIT GOT SCRAPPED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2444460 DREAMSTATION AUTO CPAP VENTILATOR, NON-CONTINUOUS (RESPIRATOR) BZD RESPIRONICS, INC. DSX500T11C

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other