FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO EVH SYSTEM

MDR report key: 2093547 · Received April 6, 2011

Report

Report Number
2242352-2011-00236
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 8, 2011
Report Date
March 8, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE INSIDE TIP OF THE COLD JAW SILICONE BOOT WAS DETACHED AND THE DETACHED PIECE WAS NOT RECEIVED. THE TOP WAS PEELING. THE HEATING ELEMENTS WERE LIFTED UP SOMEWHAT. THE JAWS WERE SOMEWHAT BURNT. THERE WAS SOME EVIDENCE OF BLOOD. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "JAW BOOT CAME OFF" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO COATING ON THE TIP CAME OFF. THERE WERE NO PATIENT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-3000 NI

Patients

Seq Age Sex Outcome Treatment
1 NA