FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2093500 · Received May 18, 2011

Report

Report Number
2531779-2011-03525
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
April 20, 2011
Report Date
April 27, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/06/2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE PUMP POWERS ON TO THE VERIFY SCREEN WITH FUNCTIONAL VIBRATORY ALARMS; THERE WERE NO AUDIBLE TONES OBSERVED DURING TESTING. INVESTIGATION REVEALED A DAMAGED CONNECTION BETWEEN THE PIEZO DISK AND THE CONTACT PAD. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

PATIENT REPORTS THAT PUMP'S SOUND IS NOT WORKING; NOTICED THIS ISSUE (B)(6). ALL TONES ARE SET TO HIGH. DURING TROUBLESHOOTING, VIBRATORY FUNCTION WAS REPORTEDLY NOT WORKING. THIS IS REPORTED DUE TO THE ALLEGATION THAT AUDIBLE AND VIBRATORY ALARM FUNCTIONS OF THE PUMP ARE NO LONGER WORKING CORRECTLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 18 YR