HS III PROXIMAL SEAL SYSTEM 3.8MM
Report
- Report Number
- 2242352-2011-00268
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Date of Event
- March 14, 2011
- Report Date
- March 14, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED SEPARATE FROM THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE ALSO RETURNED SEPARATE. THE UPPER CASING OF THE LOADING DEVICE HAD BEEN TAKEN APART AND THE SEAL HAD BEEN REMOVED. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS SLIGHT EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE REC'D CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL WOULD NOT LOAD PROPERLY" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THIS REPORTED PRODUCT LOT NUMBER. THERE WAS NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
THE HOSP REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL WOULD NOT LOAD PROPERLY. THIS OCCURRED 4 TIMES. A REPLACEMENT UNIT WAS USED TO COMPETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS . ONLY 2 OUT OF THE 4 UNITS ARE RETURNING. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HS III PROXIMAL SEAL SYSTEM 3.8MM | PROXIMAL SEAL | DXC | MAQUET CARDIOVASCULAR, LLC | HSK-3038 | 25009901 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |