FDA Adverse Event Malfunction Summary report: N

STERLING ES PTA BALLOON DILATATION CATHETER

MDR report key: 2093421 · Received May 18, 2011

Report

Report Number
2134265-2011-01851
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
April 26, 2011
Report Date
April 27, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS; THEREFORE A FAILURE ANALYSIS OF THE DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE CALCIFIED AND SEVERELY TORTUOUS LEFT EXTERNAL ILIAC ARTERY. THE 4.0X40MM STERLING BALLOON CATHETER WAS ADVANCED TO THE LESION. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 4ATMS. THE PROCEDURE WAS COMPLETED WITH A 5.0X40MM SYMMETRY BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING ES PTA BALLOON DILATATION CATHETER CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939135404010 12878861

Patients

Seq Age Sex Outcome Treatment
1 6FR TERUMO INTRODUCER SHEATH| CRUISE GUIDE WIRE