FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2093409
·
Received May 4, 2011
Report
- Report Number
- 1720753-2011-06926
- Event Type
- Malfunction
- Date Received
- May 4, 2011
- Date of Event
- April 14, 2011
- Report Date
- May 4, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE LOOSE SCREW HOLDING THE COLLIMATOR BLADE WAS TIGHTENED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM'S COLLIMATOR BLADE WOULD NOT CLOSE. NO PT INJURY WAS REPORTED.