FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO EVH SYSTEM

MDR report key: 2093402 · Received April 12, 2011

Report

Report Number
2242352-2011-00262
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 14, 2011
Report Date
March 14, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC.
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE COLD JAW SILICONE BOOT WAS PEELING ALONG THE SIDES AND TOP. THE JAWS WERE BURNT. THERE WAS SOME EVIDENCE OF BLOOD. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "JAW BOOT CRACKED" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, TWO VASOVIEW HEMOPRO JAW BOOT BROKE. THE FIRST JAW BOOT FRACTURED DURING THE PROCEDURE, ABOUT 1/2 WAY THROUGH HARVESTING. THE SECOND JAW BOOT WAS SLIDING OFF SO THE HARVESTER SLID THE JAW BOOT BACK ON AND USED IT TO COMPLETE THE PROCEDURE. NOTHING DETACHED FROM THE JAWS. THE HOSP DID NOT REPORT ANY PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC. VH-3000 25027856

Patients

Seq Age Sex Outcome Treatment
1 NA