FDA Adverse Event Malfunction Summary report: N

PULL PEG SD 20F PULL KIT

MDR report key: 2093321 · Received April 15, 2011

Report

Report Number
3006260740-2011-00108
Event Type
Malfunction
Date Received
April 15, 2011
Date of Event
March 15, 2011
Report Date
March 24, 2011
Manufacturer
C. R. BARD INC. (BASD)
Product Code
KNT
PMA / PMN Number
K915841
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS CONFIRMED. UPON RECEIPT THE INTRODUCER SHEATH REVEALED MULTIPLE TWISTED FOLDS. THE DISTAL TIP OF THE SHEATH IS JAGGED AND IRREGULAR. THE COMPLAINT INCIDENT COULD NOT BE REPLICATED IN THE LABORATORY SETTING. IT APPEARS, THE INTRODUCER SHEATH HAS BEEN DAMAGED THROUGH USE, HOWEVER, AT THIS TIME THE MECHANISM OF DAMAGE UNDETERMINED. A CHR OF LOT# HUUH1023 SHOWED NO OTHER SIMILAR PRODUCT COMPLAINT(S) FROM THIS LOT NUMBER.

Description of Event or Problem · 1

DURING THE PLACING OF THE DEVICE, THERE IS A PROBLEM DURING THE INSERTION OF THE TROCAR WHICH BEGAN PLEATED ON THE SKIN. THE DOCTOR REMOVED IT BUT THERE IS AN AIR PASSAGE IN INTRA-PERITONEAL. THE PROCEDURE WAS STOPPED BECAUSE OF A PNEUMOPERITONEUM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PULL PEG SD 20F PULL KIT KNT C. R. BARD INC. (BASD) HUUH1023

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention