FDA Adverse Event
Malfunction
Summary report: N
PULL PEG SD 20F PULL KIT
MDR report key: 2093321
·
Received April 15, 2011
Report
- Report Number
- 3006260740-2011-00108
- Event Type
- Malfunction
- Date Received
- April 15, 2011
- Date of Event
- March 15, 2011
- Report Date
- March 24, 2011
- Manufacturer
- C. R. BARD INC. (BASD)
- Product Code
- KNT
- PMA / PMN Number
- K915841
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
Narratives
Additional Manufacturer Narrative · 1
THIS COMPLAINT IS CONFIRMED. UPON RECEIPT THE INTRODUCER SHEATH REVEALED MULTIPLE TWISTED FOLDS. THE DISTAL TIP OF THE SHEATH IS JAGGED AND IRREGULAR. THE COMPLAINT INCIDENT COULD NOT BE REPLICATED IN THE LABORATORY SETTING. IT APPEARS, THE INTRODUCER SHEATH HAS BEEN DAMAGED THROUGH USE, HOWEVER, AT THIS TIME THE MECHANISM OF DAMAGE UNDETERMINED. A CHR OF LOT# HUUH1023 SHOWED NO OTHER SIMILAR PRODUCT COMPLAINT(S) FROM THIS LOT NUMBER.
Description of Event or Problem · 1
DURING THE PLACING OF THE DEVICE, THERE IS A PROBLEM DURING THE INSERTION OF THE TROCAR WHICH BEGAN PLEATED ON THE SKIN. THE DOCTOR REMOVED IT BUT THERE IS AN AIR PASSAGE IN INTRA-PERITONEAL. THE PROCEDURE WAS STOPPED BECAUSE OF A PNEUMOPERITONEUM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PULL PEG SD 20F PULL KIT | KNT | C. R. BARD INC. (BASD) | HUUH1023 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |