DA VINCI XI
Report
- Report Number
- 2955842-2024-23240
- Event Type
- Malfunction
- Date Received
- December 13, 2024
- Date of Event
- November 20, 2024
- Report Date
- November 20, 2024
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874115343
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) TO RESOLVE THE REPORTED ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE A DA VINCI PRODUCT ASSOCIATED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS (FA). HOWEVER, FA IS NOT COMPLETE.
THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) WAS PLACED ON THE SURGEON SIDE CONSOLE AND RUN IN NORMAL MODE. THE MEASUREMENT VALUE FOR HF VOLTAGE WAS TOO SMALL. THERE WERE ALSO C-34 ERRORS DURING STARTUP AND MONO COAG ACTIVATION. THE IESU ALSO HAD A COVER BENT OUTWARD ON THE SIDE LEADING TO A GAP IN THE BACK. REMOTEFE LOGS ALSO CONFIRMED C-34 AND C-38 ERRORS ON (B)(6) 2024. THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO A DEFECTIVE COMPONENT. THE IESU TRIGGERED ERROR C-34 RELATED TO LOW MEASUREMENT VALUES FOR THE HF VOLTAGE. IMAGE(S) AND/OR VIDEO CLIP(S) ASSOCIATED WITH THIS REPORTED EVENT WERE NOT SUBMITTED FOR REVIEW.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SIGMOID COLECTOMY SURGICAL PROCEDURE, THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) KEPT GIVING AN ERROR WHEN TRYING TO ACTIVATE MONOPOLAR ENERGY. THE CALLER STATED THAT THEY HAD POWER CYCLED THE IESU AND SYSTEM, BUT THE ISSUE REMAINED. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) REVIEWED THE LOGS AND CONFIRMED THE ISSUE. THE TSE CONFIRMED WITH THE CUSTOMER THAT AN EXTERNAL GENERATOR WAS NEAR THE VISION SIDE CART (VSC). THE TSE HAD THE CUSTOMER TURN OFF THE EXTERNAL GENERATOR, BUT THE ISSUE WAS NOT RESOLVED. THE TSE ALSO CONFIRMED WITH THE CUSTOMER THAT THE IESU WAS PLUGGED INTO A DEDICATED WALL POWER OUTLET. THE PROCEDURE WAS COMPLETING AS PLANNED.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1858415 | DA VINCI XI | VISION SIDE SYSTEM, RECONDITIONED | NAY | INTUITIVE SURGICAL, INC | 380721-10 | N/A | 00886874115343 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |