FDA Adverse Event Malfunction Summary report: N

DA VINCI XI

MDR report key: 20932038 · Received December 13, 2024

Report

Report Number
2955842-2024-23240
Event Type
Malfunction
Date Received
December 13, 2024
Date of Event
November 20, 2024
Report Date
November 20, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874115343
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) TO RESOLVE THE REPORTED ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE A DA VINCI PRODUCT ASSOCIATED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS (FA). HOWEVER, FA IS NOT COMPLETE.

Additional Manufacturer Narrative · 0

THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) WAS PLACED ON THE SURGEON SIDE CONSOLE AND RUN IN NORMAL MODE. THE MEASUREMENT VALUE FOR HF VOLTAGE WAS TOO SMALL. THERE WERE ALSO C-34 ERRORS DURING STARTUP AND MONO COAG ACTIVATION. THE IESU ALSO HAD A COVER BENT OUTWARD ON THE SIDE LEADING TO A GAP IN THE BACK. REMOTEFE LOGS ALSO CONFIRMED C-34 AND C-38 ERRORS ON (B)(6) 2024. THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO A DEFECTIVE COMPONENT. THE IESU TRIGGERED ERROR C-34 RELATED TO LOW MEASUREMENT VALUES FOR THE HF VOLTAGE. IMAGE(S) AND/OR VIDEO CLIP(S) ASSOCIATED WITH THIS REPORTED EVENT WERE NOT SUBMITTED FOR REVIEW.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SIGMOID COLECTOMY SURGICAL PROCEDURE, THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) KEPT GIVING AN ERROR WHEN TRYING TO ACTIVATE MONOPOLAR ENERGY. THE CALLER STATED THAT THEY HAD POWER CYCLED THE IESU AND SYSTEM, BUT THE ISSUE REMAINED. THE INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) REVIEWED THE LOGS AND CONFIRMED THE ISSUE. THE TSE CONFIRMED WITH THE CUSTOMER THAT AN EXTERNAL GENERATOR WAS NEAR THE VISION SIDE CART (VSC). THE TSE HAD THE CUSTOMER TURN OFF THE EXTERNAL GENERATOR, BUT THE ISSUE WAS NOT RESOLVED. THE TSE ALSO CONFIRMED WITH THE CUSTOMER THAT THE IESU WAS PLUGGED INTO A DEDICATED WALL POWER OUTLET. THE PROCEDURE WAS COMPLETING AS PLANNED.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1858415 DA VINCI XI VISION SIDE SYSTEM, RECONDITIONED NAY INTUITIVE SURGICAL, INC 380721-10 N/A 00886874115343

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES