FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2093106 · Received May 3, 2011

Report

Report Number
1720753-2011-06892
Event Type
Malfunction
Date Received
May 3, 2011
Date of Event
April 12, 2011
Report Date
May 3, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP REPLACED THE SURGE SUPPRESSOR PRINTED CIRCUIT BOARD. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM WOULD NOT BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1