FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2093069
·
Received May 3, 2011
Report
- Report Number
- 1720753-2011-06857
- Event Type
- Malfunction
- Date Received
- May 3, 2011
- Date of Event
- April 14, 2011
- Report Date
- May 3, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE CABLE WAS REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT TAKE PICTURES IN THE LATERAL VIEW, WOULD MAKE A TONE BUT WOULD HAVE A BLACK SCREEN. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |