FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 2092973
·
Received May 2, 2011
Report
- Report Number
- 1720753-2011-06801
- Event Type
- Malfunction
- Date Received
- May 2, 2011
- Date of Event
- April 15, 2011
- Report Date
- May 2, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE GE SERVICE REP CONDUCTED AN ONSITE INVESTIGATION. THE CPU BOARD AND THE HARD DRIVE WAS REPLACED. NO FURTHER SERVICE INFO WAS PROVIDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |