LEAD MODEL 302
Report
- Report Number
- 1644487-2011-01010
- Event Type
- Malfunction
- Date Received
- May 5, 2011
- Date of Event
- January 1, 2011
- Report Date
- April 5, 2011
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
THE INITIAL REPORT INADVERTENTLY REPORTED THE SEIZURE FREQUENCY INCORRECTLY. THE SUPPLEMENTAL REPORT #1 INADVERTENTLY REPORTED THE INCORRECT DATE. THE DATE SHOULD HAVE BEEN (B)(4) 2013.
DEVICE FAILURE IS SUSPECTED.
THE INITIAL REPORT INDICATED THAT CLINIC NOTES RECEIVED INDICATED THAT THE PATIENT WAS HAVING AN INCREASE IN SEIZURE TO 40-50 SEIZURES PER MONTH HOWEVER THERE WAS NO SEIZURE FREQUENCY LISTED ONLY THAT THE PATIENT WAS HAVING SEIZURES EVER THREE TO FOUR DAYS. FOLLOW-UP WITH THE OFFICE INDICATED THAT THE PATIENT WAS HAVING 20-40 SEIZURES PER MONTH.
ONLY A PORTION OF THE LEAD WAS RETURNED FOR ANALYSIS WHICH DID NOT REVEAL ANY ANOMALIES. DEVICE FAILURE IS SUSPECTED IN THE LEAD PORTION NOT RETURNED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
AN IMPLANT CARD LATER REPORTED THAT THE PATIENT HAD GENERATOR AND LEAD REPLACEMENT ON (B)(6) 2015. THE REASON FOR REPLACEMENT WAS DUE TO LEAD DISCONTINUITY. THE EXPLANTED PRODUCTS HAVE NOT BEEN RECEIVED BY THE MANUFACTURER TO DATE. THE RELATIONSHIP OF THE SEIZURES TO BASELINE IS UNKNOWN; HOWEVER, THE PATIENT NOTED SHE WAS NOW HAVING 40-50 SEIZURES PER MONTH.
ADDITIONAL INFORMATION AS BEE RECEIVED THAT THE PATIENT STILL HAD NOT HAD SURGERY. THE PATIENT IS DEALING WITH SO PERSONAL ISSUES AND DOES NOT WANT TO HAVE SURGERY AT THIS TIME. EVENTUAL SURGERY IS LIKELY BUT HAS NOT OCCURRED TO DATE. THE PATIENT IS CONTINUING TO HAVE 3-5 SEIZURES PER MONTH PER RECENT CLINIC NOTES.
CLINIC NOTES DATED (B)(6) 2013 WERE RECEIVED ON (B)(6) 2013. THE NOTES INDICATE THAT THE PATIENT HAS A VNS THAT HAS NOT WORKED IN THE LAST 2 YEARS, AND THE BATTERY IS DEAD. THE PATIENT CONTINUED TO HAVE 3-5 SEIZURES PER MONTH AND ARE GTC AND SMALL SEIZURES. THE PATIENT WANTED THE VNS BATTERY CHANGED; HOWEVER, IT WAS ALSO NOTED THAT THE PATIENT WANTED HER VNS REMOVED BECAUSE IT HAD BEEN INACTIVE FOR THE PAST THREE YEARS. SURGERY IS LIKELY BUT HAS NOT TAKEN PLACE.
THE EXPLANTED GENERATOR AND LEAD WERE RECEIVED BY THE MANUFACTURER FOR ANALYSIS. HOWEVER, ANALYSIS HAS NOT BEEN COMPLETED TO DATE.
ANALYSIS WAS COMPLETED ON THE DEVICES. THE ABRADED OPENING FOUND IN THE LEAD ON THE OUTER SILICONE TUBING AND THE CUT ENDS THAT WERE MADE DURING THE EXPLANTED PROCESS, MOST LIKELY PROVIDED THE LEAKAGE PATH FOR THE DRIED REMNANTS OF WHAT APPEARED TO HAVE ONCE BEEN BODY FLUIDS INSIDE THE OUTER SILICONE TUBING. FOR THE OBSERVED INNER TUBING FLUID REMNANTS, THERE WAS NO OBVIOUS PATH FOR FLUID INGRESS OTHER THAN THE CUT ENDS THAT WERE MADE DURING THE EXPLANTED PROCESS. THE CONDITION OF THE RETURNED LEAD PORTIONS IS CONSISTENT WITH CONDITIONS THAT TYPICALLY EXIST FOLLOWING AN EXPLANT PROCEDURE. NO OBVIOUS ANOMALIES WERE NOTED. CONTINUITY CHECKS OF THE RETURNED LEAD PORTIONS WERE PERFORMED, DURING THE VISUAL ANALYSIS, AND NO DISCONTINUITIES WERE IDENTIFIED. NOTE THAT SINCE THE ELECTRODE ARRAY SECTION WAS NOT RETURNED FOR ANALYSIS, AN EVALUATION AND RESULTING COMMENTARY CANNOT BE MADE ON THAT PORTION OF THE LEAD. THE GENERATOR WAS CONFIRMED TO BE AT NEAR END OF SERVICE BATTERY CONDITION. THERE WERE NO ANOMALIES FOUND WITH THE PULSE GENERATOR.
IT WAS REPORTED THAT THE PATIENT'S VNS INDICATED HIGH IMPEDANCE WITH A DCDC=7 AT A RECENT OFFICE VISIT. CLINIC NOTES RECEIVED INDICATE THAT THE PATIENT'S SEIZURE FREQUENCY HAS INCREASED RECENTLY TO APPROXIMATELY 40-50 SEIZURES PER MONTH, HOWEVER, THE RELATIONSHIP TO PRE-VNS BASELINE IS UNKNOWN. THE PATIENT HAS NOT FELT STIMULATION FOR APPROXIMATELY 3 MONTHS. THE NOTES ALSO STATED THAT THE VNS GENERATOR RECENTLY "STOPPED WORKING" AND SHOWED "AN ERROR SAYING THAT MOST LIKELY THE BATTERY IS DEAD." A BATTERY LIFE CALCULATION ESTIMATED 2.71 YEARS UNTIL END OF SERVICE. FOLLOW-UP WITH THE PHYSICIAN'S OFFICE FOUND THAT THE PATIENT'S VNS HAS NOT BEEN DISABLED BUT SHE IS PLANNED TO BE DISABLED AT THE NEXT VISIT. CHEST AND NECK X-RAYS ARE ALSO PLANNED FOR THE NEXT VISIT TO CHECK FOR A LEAD FRACTURE. THE PATIENT DID NOT ATTEND HER LAST VISIT WHEN THEY WERE PLANNING TO TAKE X-RAYS AND DISABLE THE VNS AND THE SITE HAS NOT BEEN ABLE TO CONTACT THE PATIENT. NO TRAUMA OR MANIPULATION HAS BEEN REPORTED. SURGERY TO REPLACE THE VNS LEAD AND GENERATOR WILL BE PLANNED FOLLOWING THE NEXT VISIT WITH THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LEAD MODEL 302 | LYJ | CYBERONICS, INC. | 302-20 | 1421 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Required Intervention |