FDA Adverse Event
Malfunction
Summary report: N
7900
MDR report key: 2092824
·
Received May 5, 2011
Report
- Report Number
- 9680959-2011-01182
- Event Type
- Malfunction
- Date Received
- May 5, 2011
- Date of Event
- April 5, 2011
- Report Date
- May 5, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE CABLE TO THE GENERATOR WAS REPAIRED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CANBUS ERROR MESSAGE, AND WOULD NOT PRODUCE AN X-RAY. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 7900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS GMBH | 7900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |