FDA Adverse Event
Malfunction
Summary report: N
9600
MDR report key: 2092727
·
Received May 5, 2011
Report
- Report Number
- 1720753-2011-06945
- Event Type
- Malfunction
- Date Received
- May 5, 2011
- Date of Event
- April 13, 2011
- Report Date
- May 5, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE GE SERVICE REPRESENTATIVE CONDUCTED AN ONSITE INVESTIGATION. THE SRAM AND THE TECHNICAL PROCESSOR WERE REPLACED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CHECKSUM ERROR MESSAGE. THIS MALFUNCTION OCCURRED INSIDE A PROCEDURE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |