FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX WINKELSTUCK 25LP
MDR report key: 2092617
·
Received May 9, 2011
Report
- Report Number
- 3003637274-2011-00028
- Event Type
- Injury
- Date Received
- May 9, 2011
- Date of Event
- April 1, 2011
- Report Date
- April 12, 2011
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
DURING A STANDARD TREATMENT, A DENTAL ELECTRICAL HANDPIECE GOT HOT AND BURNED PATIENT ON CHEEK. THE ANALYSIS DID SHOW THAT THE HEAD HAD A DENT WHICH CAUSED THE BACKCAP BUTTON TO STICK IN. THIS CAUSED A LOT OF FRICTION AND AS A RESULT THE HEAD HEATED UP. ISSUE OCCURRED IN (B)(6) WITH SIMILAR DEVICE THAN DISTRIBUTED IN THE US.
Description of Event or Problem · 1
DURING A STANDARD TREATMENT, A DENTAL ELECTRICAL HANDPIECE GOT HOT AND BURNED PATIENT ON CHEEK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX WINKELSTUCK 25LP | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |