FDA Adverse Event Injury Summary report: N

GENTLEPOWER LUX WINKELSTUCK 25LP

MDR report key: 2092617 · Received May 9, 2011

Report

Report Number
3003637274-2011-00028
Event Type
Injury
Date Received
May 9, 2011
Date of Event
April 1, 2011
Report Date
April 12, 2011
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

DURING A STANDARD TREATMENT, A DENTAL ELECTRICAL HANDPIECE GOT HOT AND BURNED PATIENT ON CHEEK. THE ANALYSIS DID SHOW THAT THE HEAD HAD A DENT WHICH CAUSED THE BACKCAP BUTTON TO STICK IN. THIS CAUSED A LOT OF FRICTION AND AS A RESULT THE HEAD HEATED UP. ISSUE OCCURRED IN (B)(6) WITH SIMILAR DEVICE THAN DISTRIBUTED IN THE US.

Description of Event or Problem · 1

DURING A STANDARD TREATMENT, A DENTAL ELECTRICAL HANDPIECE GOT HOT AND BURNED PATIENT ON CHEEK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX WINKELSTUCK 25LP DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LP

Patients

Seq Age Sex Outcome Treatment
1 Other