FDA Adverse Event Malfunction Summary report: N

BD SYNAPSYS¿

MDR report key: 20925741 · Received December 13, 2024

Report

Report Number
1119779-2024-00993
Event Type
Malfunction
Date Received
December 13, 2024
Date of Event
November 22, 2024
Report Date
December 3, 2024
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JQP
UDI-DI
00382904441500
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1. INITIAL ADDRESS: (B)(6). H6. INVESTIGATION SUMMARY : THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION OF A COMPLAINT INVOLVING SYNAPSYS. IT WAS REPORTED THAT THE SYNAPSYS WAS NOT COMMUNICATING WITH BACTEC. NO OTHER ISSUES WERE REPORTED. BD INVESTIGATED THE ISSUE. SYNAPSYS ONLY HAD A PARTIAL ACCESSION NUMBER. THE ACCESSION NUMBER IN SYNAPSYS DID NOT MATCH WHAT WAS IN THE FX. THIS WAS CORRECTED BY ASSOCIATING THE CORRECTION ACCESSION NUMBER. THE MESSAGES WERE BACKED UP IN SYNAPSYS. THE SYNAPSYS BACKLOGGED PAYLOADS WERE PROCESSED. THE SITE HAD PROBLEMS OVER THE WEEKEND. THE DATA PROCESSOR WAS RESTARTED, AND STUCK MESSAGES WERE TRANSFERRED. THE ISSUE IS RESOLVED. THIS IS A CONFIRMED FAILURE WITH A BD PRODUCT. REVIEW FOUND COMPLAINTS OF THIS TYPE WERE BELOW STATISTICAL CONTROL FOR THE MONTH OF NOVEMBER. DEVICE HISTORY RECORD REVIEW WAS NOT APPLICABLE AS THIS IS A STANDALONE SOFTWARE PRODUCT AND THE OPERATION/FUNCTIONALITY OF THIS TYPE OF PRODUCT IS CONFIRMED AT THE TIME OF INSTALLATION. REPORTS OF THIS TYPE WILL CONTINUE TO BE MONITORED. IF YOU HAVE ANY ADDITIONAL QUESTIONS OR CONCERNS, PLEASE DO NOT HESITATE TO CONTACT BD TECHNICAL SUPPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD SYNAPSYS¿ AN UNSPECIFIED NUMBER OF PARTIAL ACCESSIONS WERE ASSOCIATED TO THE SEQUENCE NUMBERS, DESPITE THE FULL ACCESSION NUMBER PRESENTED IN THE INSTRUMENT. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1936110 BD SYNAPSYS¿ NA JQP BECTON DICKINSON & CO. (SPARKS) 00382904441500

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown