FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2092474
·
Received May 2, 2011
Report
- Report Number
- 1720753-2011-06835
- Event Type
- Malfunction
- Date Received
- May 2, 2011
- Date of Event
- April 13, 2011
- Report Date
- May 2, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE FLUOROSCOPY BEAM WAS CHECKED IN THE AP AND LATERAL POSITIONS, AND RADIATION OUTPUT WAS WITHIN SPECIFICATIONS. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT PERFORM FLUOROSCOPIC X-RAY IN THE LATERAL POSITION. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |