FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2092447 · Received May 2, 2011

Report

Report Number
1720753-2011-06814
Event Type
Malfunction
Date Received
May 2, 2011
Date of Event
April 15, 2011
Report Date
May 2, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION AND REPLACED THE CAMERA ASSEMBLY AND PERFORMED A BEAM ALIGNMENT. THE INTERCONNECT CABLE WAS REPLACED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A CASE AND DURING FLUOROSCOPY, THE SYSTEM'S MONITOR SCREEN WENT DARK. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1