FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2092273 · Received April 27, 2011

Report

Report Number
1720753-2011-06676
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
April 5, 2011
Report Date
April 27, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ON-SITE INVESTIGATION. THE COLLIMATOR WAS REPLACED AND CALIBRATED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED. THIS COULD BE A POSSIBLE ACCIDENTAL RADIATION OCCURENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9600 SYSTEM WOULD NOT EXPOSE AN X-RAY AND DISPLAYED A BAD IRIS POTENTIOMETER ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1