FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2092227 · Received April 27, 2011

Report

Report Number
1720753-2011-06643
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
April 5, 2011
Report Date
April 27, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MFR'S SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE SERVICE REP REPLACED THE HIGH VOLTAGE CABLE, THE BRAKE HANDLE, THE KEYPAD EXTENSION, SINGLE BOARD COMPUTER CIRCUIT BOARD, AND THE SINGLE BOARD COMPUTER COIN BATTERY. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9800 SYSTEM MONITORS WOULD NOT TURN ON AND FRAYED CABLES WERE FOUND. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1