- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MFR'S SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE SNUBBER PRINTED CIRCUIT BOARD WAS REPLACED. THE HIGH VOLTAGE ADJUSTMENTS AND FILAMENT CALIBRATIONS WERE MADE. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9800 SYSTEM WOULD NOT DISPLAY A LIVE IMAGE. NO PT INJURY WAS REPORTED.