FDA Adverse Event Malfunction Summary report: N

CIC PRO CLINICAL INFORMATION CENTER

MDR report key: 2092187 · Received April 27, 2011

Report

Report Number
2124823-2011-00048
Event Type
Malfunction
Date Received
April 27, 2011
Date of Event
March 29, 2011
Report Date
April 27, 2011
Manufacturer
GE HEALTHCARE
Product Code
DSI
PMA / PMN Number
K062976
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

A SITE REPORTED A LOSS OF TELEMETRY MONITORING FOR 5 PTS FOR APPROX 2 HRS FOLLOWING A POWER OUTAGE. WHEN POWER WAS RESTORED THE CENTRAL MONITORING SYSTEM DID NOT RE-ESTABLISH COMMUNICATION WITH THE TELEMETRY MONITORING UNITS. NO ALTERNATE MONITORING WAS ESTABLISHED DURING THIS TIME. MONITORING WAS RESTORED BY COMPLETING A FULL POWER DOWN AND REBOOT OF THE SERVER. THERE WERE NO PT INJURIES REPORTED. GE HEALTHCARE'S INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A F/U REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION HAS BEEN COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CIC PRO CLINICAL INFORMATION CENTER CENTRAL MONITORING SYSTEM DSI GE HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1