STEALTHSTATION TREON TREATMENT GUIDANCE SYSTEM
MDR report key: 2092172
·
Received April 27, 2011
Report
- Report Number
- 1723170-2011-00912
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- January 31, 2011
- Report Date
- January 31, 2011
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- PMA / PMN Number
- K050438
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. DEVICE MANUFACTURE DATE NOT AVAILABLE AT TIME OF THIS REPORT. THE ISO C/STEALTHSTATION INTERFACE WAS SUCCESSFULLY RE-CALIBRATED AND TESTED PERFECTLY ACCURATE.
Description of Event or Problem · 1
A MEDTRONIC REPRESENTATIVE REPORTED THAT DURING PLANNED MAINTENANCE FOR THE SITE'S STEALTHSTATION TREON, THE INTERFACE BETWEEN THE ISO C, (A NON-MEDTRONIC PRODUCT), AND THE STEALTHSTATION TREON WAS FOUND TO BE OUT OF CALIBRATION. THE CALIBRATION WAS INACCURATE BY 7MM LATERAL AND ANTERIOR. RE-CALIBRATION IS REQUIRED. THERE WAS NO PATIENT PRESENT. THE SYSTEM WAS NEVER USED IN A CASE WHEN IT WAS OUT OF CALIBRATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STEALTHSTATION TREON TREATMENT GUIDANCE SYSTEM | STEREOTAXIC INSTRUMENT (HAW) | HAW | MEDTRONIC NAVIGATION, INC. | TREON | NA |