FDA Adverse Event
Malfunction
Summary report: N
INFINITI VISION SYSTEM OZIL
MDR report key: 2092167
·
Received April 27, 2011
Report
- Report Number
- 2028159-2011-00457
- Event Type
- Malfunction
- Date Received
- April 27, 2011
- Date of Event
- March 28, 2011
- Report Date
- March 28, 2011
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REP EXAMINED THE SYSTEM AND OBSERVED HANDPIECE NOT WORKING AS PER SPECIFICATIONS. THE SYSTEM WAS THEN TESTED AND MET PRODUCT SPECIFICATIONS. NO PARTS WERE REPLACED. ADD'L INFO REGARDING PRODUCT EVAL IS PENDING. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WHEN ADD'L REPORTABLE INFO BECOMES AVAILABLE. (B)(4).
Description of Event or Problem · 1
A NURSE REPORTED THE SYSTEM WOULD NOT RECOGNIZE THE HANDPIECE DURING SURGERY. THE SYSTEM WAS SWITCHED OUT AND THE CASE WAS COMPLETED. THERE WAS NO PT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | INFINITI OZIL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL TORSIONAL HANDPIECE |