FDA Adverse Event Death Summary report: N

CONSULTA CRT-D

MDR report key: 2092058 · Received May 18, 2011

Report

Report Number
2647346-2011-00704
Event Type
Death
Date Received
May 18, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND PRIMARY ANALYSIS REVEALED NO ANOMALIES FOUND. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND PRIMARY ANALYSIS REVEALED NO ANOMALIES FOUND. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND PRIMARY ANALYSIS REVEALED NO ANOMALIES FOUND.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND PRIMARY ANALYSIS REVEALED NO ANOMALIES FOUND. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND PRIMARY ANALYSIS REVEALED NO ANOMALIES FOUND. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND PRIMARY ANALYSIS REVEALED NO ANOMALIES FOUND.

Description of Event or Problem · 1

THE IMPLANTABLE CARDIAC DEFIBRILLATOR WAS RETURNED WITH NO INFORMATION. A DATABASE SEARCH REVEALED THAT THE PATIENT HAD EXPIRED APPROXIMATELY 7 MONTHS AFTER IMPLANT. FOLLOW UP HAS BEEN INCONCLUSIVE, CAUSE OF DEATH WAS REQUESTED AND NOT YET RECIEVED. NO COMPLAINTS OR ALLEGATIONS HAVE BEEN MADE AGAINST THE SYSTEM OR ANY OF THE INDIVIDUAL COMPONENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK ASKU

Patients

Seq Age Sex Outcome Treatment
1 72 YR Death| O