FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2092032 · Received May 18, 2011

Report

Report Number
1423500-2011-06127
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
April 22, 2011
Report Date
April 22, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT OF A SYSTEM ERROR 2240 ALARM WAS CONFIRMED. PER INFORMATION PROVIDED BY THE PATIENT THE CAUSE WAS DETERMINED TO BE USE ERROR (THE PATIENT STATED THAT THEY WERE 20 MINUTES LATE IN RECONNECTING TO THE INSTRUMENT). THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED BY THE CUSTOMER. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DRAIN 2 OF 3. THE HOME PATIENT (HP) STATED THAT HE DISCONNECTED DURING DWELL THEN GOT BACK 20 MINUTES LATE TO THE HC. THE HP RECONNECTED BUT THE HC WAS ALARMING. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM. THE TSR HAD THE HP CLOSED THE CLAMPS AND CYCLE POWER TO CLEAR THE ALARM. THE TSR ADVISED THE HP TO DISCARD THE SUPPLIES AND FINISH WITH MANUAL. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT. DURING FOLLOW UP, THE HP STATED HE IS STILL USING THE SPIKE CONNECTORS AND THAT HE WAS STILL DOING THERAPY. NO FURTHER INFORMATION WAS GIVEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 50 YR HOMECHOICE