FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2092021 · Received May 18, 2011

Report

Report Number
2531779-2011-03508
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
April 18, 2011
Report Date
April 19, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6). THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED THAT ALL BUTTONS OTHER THAN THE AUDIO BOLUS BUTTON INTERMITTENTLY ACTIVATED PUMP FUNCTIONS WHEN PRESSED. THE AUDIO BOLUS BUTTON FUNCTIONED PROPERLY. CONTAMINATION WAS OBSERVED UNDER ALL KEY CONTACTS EXCEPT THE AUDIO BOLUS BUTTON CONTACT. UNRELATED TO THE COMPLAINT AN UNIDENTIFIED DISPLAY FAILURE WAS OBSERVED.

Description of Event or Problem · 1

THE DISTRIBUTOR REPORTED THAT THE CONTRAST BUTTON WAS STICKING AND ALL THE WRITING ON THE BUTTONS WORE OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1