FDA Adverse Event Death Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2092014 · Received May 18, 2011

Report

Report Number
2649622-2011-07573
Event Type
Death
Date Received
May 18, 2011
Date of Event
April 8, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. A POWER ON RESET OCCURRED (B)(6) 2011, WHICH IS AFTER THE NOTED DATE OF DEATH. NO DATA IS AVAILABLE BEFORE THE DATE OF DEATH WHICH IS NOTED AS (B)(6) 2011. (B)(4) NO ANOMALIES FOUND, SEVERAL CONDUCTORS BLOOD/BODY FLUID (NOT OBSTRUCTED), OUTER INSULATION COSMETIC ESC AND DEPRESSION, OUTER TUBING OVERLAY BREACHED CUT, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION COSMETIC CUT, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION COSMETIC DEPRESSION AND BREACHED CUT, DAMAGED AT IMPLANT. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. A POWER ON RESET OCCURRED (B)(6) 2011, WHICH IS AFTER THE NOTED DATE OF DEATH. NO DATA IS AVAILABLE BEFORE THE DATE OF DEATH WHICH IS NOTED AS (B)(6) 2011.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND A POWER ON RESET OCCURRED AND IT NOTED TO HAVE OCCURRED AFTER THE DATE OF DEATH. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. A POWER ON RESET OCCURRED (B)(4) 2011, WHICH IS AFTER THE NOTED DATE OF DEATH. NO DATA IS AVAILABLE BEFORE THE DATE OF DEATH WHICH IS NOTED AS (B)(4) 2011. (B)(4) NO ANOMALIES FOUND, SEVERAL CONDUCTORS BLOOD/BODY FLUID (NOT OBSTRUCTED), OUTER INSULATION COSMETIC ESC AND DEPRESSION, OUTER TUBING OVERLAY BREACHED CUT, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION COSMETIC CUT, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION COSMETIC DEPRESSION AND BREACHED CUT, DAMAGED AT IMPLANT. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Description of Event or Problem · 1

ASKU

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS DECEASED AND DIED AT HOME UNEXPECTEDLY. IT WAS FURTHER REPORTED THAT DURING A POST-MORTEM INTERROGATION OF THE DEVICE, AN ELECTRICAL RESET WAS NOTED TO HAVE OCCURRED. UPON FURTHER REVIEW, THE ELECTRICAL RESET OCCURRED ONE WEEK POST-MORTEM. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Death VEDR01 IMPLANTABLE PULSE GENERATOR| VEDR01 IMPLANTABLE PULSE GENERATOR| VEDR01 IMPLANTABLE PULSE GENERATOR