FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC APPLIER

MDR report key: 2091921 · Received May 18, 2011

Report

Report Number
3005075853-2011-02022
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
April 12, 2011
Report Date
April 13, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED SIXTEEN CLIPS AS INTENDED AND ONE SCISSOR CLIP CLASS III. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE MALFORMED CLIP. PLEASE NOTE THAT THIS FINDING IS UNRELATED WITH THE INCIDENT REPORTED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE CONTAINED ONLY THREE CLIPS. THERE WERE NO PATIENT CONSEQUENCES. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGACLIP** ENDOSCOPIC APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA G4UD95

Patients

Seq Age Sex Outcome Treatment
1