ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2091853 · Received May 17, 2011

Report

Report Number
2531779-2011-03504
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 25, 2011
Report Date
April 26, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTONS WERE FOUND TO BE RESPONDING APPROPRIATELY TO PRESSES. THE KEYPAD WAS REMOVED AND NO DEFECTS WERE FOUND WITH THE BUTTON CONTACTS. UNRELATED TO THE COMPLAINT, THE DISPLAY LENS WAS FOUND TO LEAK; NO EVIDENCE OF MOISTURE WAS OBSERVED INSIDE THE PUMP.

Description of Event or Problem · 1

PATIENT'S MOTHER REPORTS PUMP LOCKED UP ON VERIFY SCREEN AFTER EXPOSURE TO POOL WATER AT AMUSEMENT PARK. SHE ALSO REPORTS CONTRAST BUTTON IS WORKING, HOWEVER, UP AND DOWN BUTTONS ARE NOT RESPONDING. PATIENT DOES NOT CLEAN PUMP. PATIENT WEARS PUMP IN POCKET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 8 YR