ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2091827
·
Received May 17, 2011
Report
- Report Number
- 2531779-2011-03502
- Event Type
- Malfunction
- Date Received
- May 17, 2011
- Report Date
- April 26, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(6). THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE OK AND DOWN KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.
Description of Event or Problem · 1
PATIENT'S MOTHER REPORTS THAT OK AND DOWN ARROW BUTTONS REQUIRES MULTIPLE PRESSES TO RESPOND. PATIENT'S MOTHER REPORTS NOTICED ISSUE A MONTH AGO AND HAS BEEN GETTING WORSE. ALL OTHER BUTTONS FEEL NORMAL AND RESPOND NORMALLY TO PRESSES. PATIENT DOES NOT CLEAN PUMP. PATIENT WEARS PUMP IN POCKET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |