FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2091825 · Received May 17, 2011

Report

Report Number
1423500-2011-06090
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
May 2, 2011
Report Date
May 2, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) OCCURRED DURING FILL WAS NOT CONFIRMED; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE THE SUPPLY BAG CAME DISCONNECTED FROM THE SUPPLY LINE. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. THIS REVIEW FINDS THE LABELING ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE LOT NUMBER WAS NOT PROVIDED; THEREFORE A BATCH REVIEW CANNOT BE CONDUCTED. THE SAMPLE WAS DISCARDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

A HOME PATIENT (HP) CONTACTED (B)(4) REGARDING A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOME CHOICE (HC) UNIT DURING FILL. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP WITH CLEARING THE ALARM THEN EXPLAINED A SE 2240 INDICATES A LARGE AMOUNT OF AIR HAS ENTERED SETUP AND THAT THE HP NEEDED TO END THERAPY AND CONTACT THEIR NURSE. THE PATIENT WAS INVOLVED, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. A FOLLOW UP WAS MADE VIA PHONE CALL. PER THE CARE GIVER (CG), THEY STARTED OVER WITH NEW SUPPLIES AND THE HP RESUMED THERAPY WITHOUT ANY PROBLEMS. THE HP FOLLOWED UP WITH THE NURSE REGARDING THE ALARM RECEIVED. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 73 YR HOMECHOICE