INJECTION GOLD PROBE
Report
- Report Number
- 3005099803-2011-01708
- Event Type
- Malfunction
- Date Received
- May 17, 2011
- Date of Event
- April 12, 2011
- Report Date
- April 21, 2011
- Manufacturer
- BOSTON SCIENTIFIC - COSTA RICA
- Product Code
- KNS
- PMA / PMN Number
- K961349
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE COMPLAINANT WAS UNABLE TO PROVIDE THE SUSPECT DEVICE LOT NUMBER; THEREFORE, THE DEVICE EXPIRATION AND MANUFACTURE DATES ARE UNKNOWN. ACCORDING TO THE COMPLAINANT, THE SUSPECT DEVICE HAS BEEN DISPOSED OF AND IS NOT AVAILABLE FOR RETURN. A DEVICE EVALUATION CANNOT BE PERFORMED. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH REPORT WILL BE FILED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN INJECTION GOLD PROBE DEVICE WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE THE NEEDLE BECAME STUCK IN THE EXTENDED POSITION; IT WOULD NOT RETRACT. THE PHYSICIAN HAD ALREADY COMPLETED THE PROCEDURE WITH THIS DEVICE WHEN THE ISSUE OCCURRED. THE EXTENDED NEEDLE DAMAGED THE SCOPE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. AT THE CONCLUSION OF THE PROCEDURE, THE PATIENT'S CONDITION WAS REPORTED TO BE FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INJECTION GOLD PROBE | UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) | KNS | BOSTON SCIENTIFIC - COSTA RICA | M00560150 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |