FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2091773 · Received May 17, 2011

Report

Report Number
2531779-2011-03490
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 19, 2011
Report Date
April 27, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1, DATE OF SUBMISSION (B)(4) 2011-DEVICE EVALUATION: THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE TORN AT THE UP AND DOWN ARROW BUTTONS. A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP WAS PERFORMED AND THE PUMP WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE FIRMWARE REVISION AND THE MANUFACTURE DATE ARE (D1E, 2010/08). THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE DISTRIBUTOR CLAIMED THAT THE UP/DOWN/OK BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1