FLOGARD
Report
- Report Number
- 6000001-2011-04145
- Event Type
- Malfunction
- Date Received
- May 17, 2011
- Date of Event
- May 2, 2011
- Report Date
- May 2, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE REPORTED CONDITION OF BATTERY LOW ALARM WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS A DEPLETED MAIN BATTERY. THE MAIN BATTERY WAS REPLACED TO CORRECT THE REPORTED CONDITION. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. (B)(4). HOWEVER, BAXTER WILL CONTINUE TO MONITOR AND REPORT ON DEATH AND SERIOUS INJURY (DSI) EVENTS TO ASSESS THE IMPACT ON PATIENT SAFETY.
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND WITH A BATTERY LOW ALARM. THIS CONDITION INTERRUPTED DELIVERY AS THE PUMP WAS BEING TESTED. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |