FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2091706 · Received May 17, 2011

Report

Report Number
6000001-2011-04136
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 28, 2011
Report Date
April 28, 2011
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K883577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED MALFUNCTION OF "A FALSE OCCLUSION ALARM" WAS CONFIRMED AND REPRODUCED. THE ROOT CAUSE WAS ATTRIBUTED TO AN OCCLUSION SWITCH OUT OF ADJUSTMENT. THE OCCLUSION SWITCH WAS ADJUSTED TO FIX THE PROBLEM. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. A DEVICE HISTORY REVIEW WAS PERFORMED FINDING THAT THE DATA CARD HAS BEEN DISCARDED PER RETENTION PERIOD DOCUMENTED ON 20J. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT AN INFUSOR PUMP IN WHICH THERE WAS A FALSE OCCLUSION ALARM. THE EVENT OCCURRED IN THE OPERATING ROOM. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION

Patients

Seq Age Sex Outcome Treatment
1