FDA Adverse Event Malfunction Summary report: N

LONG 45 ENDOCUTTER

MDR report key: 2091700 · Received May 17, 2011

Report

Report Number
3005075853-2011-02007
Event Type
Malfunction
Date Received
May 17, 2011
Report Date
April 21, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INCOMPLETE/INTERRUPTED FIRING THE ANALYSIS RESULTS FOUND THAT THE LONG45A DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH TWO RELOADS PRESENT. RELOAD B WAS RECEIVED FULLY FIRED; RELOAD C WAS RECEIVED PARTIALLY FIRED WHICH INDICATES THAT THE DEVICE'S FIRING CYCLE WAS INTERRUPTED. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT, AND ALL THE STAPLES FORMED AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE CARTRIDGE WAS LOADED INTO THE DEVICE WITHOUT DIFFICULTIES AND DID NOT FELL OUT DURING THE VISUAL OR FUNCTIONAL TESTING. ALTHOUGH NO CONCLUSION COULD BE REACH ON WHAT CAUSED THE REPORTED EVENT IT IS POSSIBLE THAT THE RELOAD WAS LOADED INCORRECTLY. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE MAGAZINE SLIPPED OUT OF THE INSTRUMENT DURING CLOSING. NO FURTHER INFORMATION IS AVAILABLE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LONG 45 ENDOCUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK

Patients

Seq Age Sex Outcome Treatment
1 6R45B, LOT#- H43364